Appropriate Location and Equipment
Authors: Sciard, Didier A.; Matuszczak, Maria E.
Title: Landmarks for Peripheral Nerve Blocks: Upper and Lower Extremities, 2nd Edition
Copyright ©2008 Lippincott Williams & Wilkins
> Front of Book > Principles of Neurostimulation
Principles of Neurostimulation
Appropriate Location and Equipment
-
Room equipped with standard ASA monitors
-
Immediate access to resuscitative drugs and equipment
-
Equipment checked
-
Neurostimulators able to deliver a
current of up to 5-mA intensity, frequency of 1 or 2 Hz (1 or 2
impulses per second), impulse duration of 0.1 to 1 msec (0.1 to 0.3
msec for a motor nerve stimulation and 1 msec for a sensory nerve
stimulation) -
Insulated needles of 1, 2, 4, and 6 inches (2.5, 5, 10, and 15 cm)
-
Catheter sets with 2-, 4-, and 6-inch (5-, 10-, and 15-cm) introducer needles
P.viii
Technique
-
Check nerve stimulator and connecting cables for their proper function.
-
Skin prep and local infiltrate for needle insertion.
-
Minimal distance between skin electrode and needle.
-
After skin puncture, gradually increase the intensity up to 1.5 mA at 0.3 msec.
-
Search in 3 axes (see picture on page x).
-
As soon as the first twitch is obtained,
impulse duration (msec) should be decreased to 0.1 msec. Then intensity
(mA) is slowly decreased while approaching the nerve to a minimal
stimulation intensity of 0.5 to 0.3 mA. -
Aspirate for blood.
-
Loss of motor response after injection of 1 mL of local anesthetic and return of the motor response when intensity is increased.
-
No resistance or pain during injection.
-
Slow and fragmented injection.
P.x